对露点传感器和仪表进行校验的集成系统
露点校验系统 产品详情
product featuresself-contained systemwide measurement range-100 to +20°c technical specification
dcs80
range -80 to +20°c dew point
air dryer michell psd-2 dryer
generator method dg-5 with solenoid controlled needle valves
reference s4000 rs cooled mirror dewpointmeter
calibration flow rate 2 to 5 nl/min;*
dcs100
range -100 to +20°c dew point
air dryer michell psd-4 super dryer
generator method vds system with mass flow controllers
reference s4000 trs cooled mirror dewpointmeter
calibration flow rate 5 nl/min;**
dcs80 and dcs100
best system uncertainty ±0.2°c dew point (k = 2) at +20°c dew point
set-point precision ±0.5 °c dew point
carrier gas oil-free compressed air (compressor supplied)
operating temperature +15 to +30°c
traceability directly to npl and nist
power 220/240 v or 100/130 v, 50/60 hz
housing wheeled 19 rack system, 1.9m high
weight 120kg (approx)
for a more detailed specification, please consult michell instruments´ technical sales department
*dependent upon generated dew point
**fixed flow
reference instrumentation calibrated to nist and npl standards