Basic Info.
LCD Display
With LCD Display
Ultra High Temperature Alarm
Ultra High Temperature Alarm
Sterilization Information Recording and Printing
Sterilization Information Recording and Printing
Transport Package
Paper Carton
Specification
690*485*455
Production Capacity
1000 Sets
Product Description
Pulse vacuum class B benchtop steam sterilzer 1. Pulse vacuum exhaust air trapped in the chamber, boostering sterilization process.
2. Built-in printer records temperature, time, pressure of vacuum processes, sterilization process and other info., so you can trace every sterilization cycle when necessary.
3. Concise intuitive control board where you can select sterilization process, monitor temperature, pressure, water volume, door status etc.
4. Preset sterilization cycles for your selection
Four preset programs parameters:
No. | Programs | Recommended Load Types |
Packing | Temperature/°C | Pressure/bar | Sterilization Time/min | Drying Time/min |
1 | Unwrapped | 134 | 2.1 | 4 | 10 | Unwrapped solid loads |
2 | Unwrapped | 121 | 1.1 | 20 | 10 |
3 | Wrapped | 134 | 2.1 | 4 | 30 | Unwrapped textile load Single layer textile load Double-layer textile loads Single-layer hollow loads Double-layer solid or hollow loads |
4 | Wrapped | 121 | 1.1 | 20 | 30 |
5 | Vacuum Test | - | - | - | | - |
5. Specifications for different models of sterilizers
Chamber volume | Max. Instrument Load/kg | Max. Textile Load/kg | Max. Single Load Weight/kg | Max. Load of Each Tray/kg |
8L | 1.5 | 0.4 | 0.4 | 0.5 |
12L | 2.5 | 0.6 | 0.6 | 0.8 |
18L | 3.5 | 1 | 0.8 | 0.7 |
23L | 5 | 1.2 | 1.2 | 1 |
Dimensions for models
Volume | L(mm) | W(mm) | H(mm) |
8L | 580±10 | 420±10 | 385±10 |
12L | 580±10 | 420±10 | 385±10 |
18L | 690±10 | 485±10 | 455±10 |
23L | 690±10 | 485±10 | 455±10 |