The Totalflow NGC8206 has an extensive built-in list of tests which are performed
each time the unit is started. This start-up testing may be disabled, but Totalflow
recommends that it be left enabled. These diagnostics consist of four areas of
testing:
• Carrier Pressure Regulator Test
• Oven Temperature Test
• Processor Control Test
• Stream Test
These start-up tests may also be performed on a regular schedule. Please see the
PCCU Help files for more information on scheduling diagnostics.
1.11.1 Carrier Pressure Regulator Tests
This test compares the actual column pressure to the column pressure set point
using carrier gas. A failure of this test indicates that the carrier pressure is not
meeting or over exceeding the expected level of pressure.
1.11.2 Oven Temperature Test
This test compares the actual oven temperature to the oven temperature set point.
A failure of this test indicates that the oven is not maintaining the required
temperature.
1.11.3 Processor Control Test
This test contains three test areas: column 1 carrier pressure, column 2 carrier
pressure and oven temperature. In each area, the test measures the effort
required to maintain the required value. From those measurements, the test
develops a standard deviation and makes a comparison. The failure of any of
these comparisons indicates an erratic deviation exists, meaning the processor is
not able to control the function.